BioCommand® Software
Sample Program Screen

BioCommand's powerful programming feature has been designed to help scientists and process technicians write sophisticated control strategies to automate setpoint changes based on culture conditions and time. No programming skills needed. Simply drag and drop the logic blocks into place to create “If-Then” statements. In the first example above, if the DO is greater than 40, the software automatically adjusts the setpoint of Pump 1 to 100. If the statement is false, the setpoint is set to 0. Programs can also be used to calculate values which are not already present as loops based on already-available information, as shown in the second example. Here, tip speed is calculated from a mathematical formula and agitator speed. This value becomes a new loop and can be recorded as part of the batch information. For advanced programmers, Visual Basic script can also be used.
Available in Batch Control and Batch Control Plus only.